TS7030 is 3 nh using autonomous core technology research and development of A portable light color difference instrument, is the high level in spectral architecture color difference instrument, in addition to ensure accurate relative Δ E at the same time, also to ensure the accuracy of the absolute value of L, A and B for A long time, anytime, anywhere can pass the international standards and national standards of measurement. Using built-in silicon photodiode array (double row group 32) sensors, import the whiteboard, repetitive Δ E * ab easily control within 0.06, both measurement speed and convenience of the operation at the same time, connecting the PC software or even the PC software can quickly judging color difference measurement, powerful, 8 mm diameter meet plastic electronics, paint and ink, textile printing and dyeing, printing, such as the ceramic industry production and the quality of accurate color control.
Spectrophotometer TS7030 product features
1. Φ 8 mm diameter measurement, 10 ° Angle, observer can meet most of the industry customer requirements;
2. Combination of full-spectrum LED light source, plane grating and true color capacitive touch screen, with a high starting point;
3. Silicon photodiode array (32 groups with double rows) sensor, display precision 0.01, good repeatability;
4. The light source life is more than 3 million measurements in 5 years, and the lithium battery takes 6,000 measurements within 8 hours, which can be recharged repeatedly and has a long battery life;
5. 1000 data storage standard samples and 20000 samples were used for quick data search, analysis and comparison.
Model | TS7030 |
Optical Geometry | D/8(diffused illumination, 8-degree viewing angle) |
SCI/SCE Mode | |
Comply to CIE No.15,GB/T 3978,GB 2893,GB/T 18833,ISO7724-1,ASTM E1164,DIN5033 Teil7 | |
Characteristic | single apertures, more adaptability; Used for accurate color measurement and quality control in plastic electronics, paint and ink, textile and garment printing and dyeing, printing, ceramics and other industries |
Integrating Sphere Size | Φ40mm |
Light Source | Combined full spectrum LED light source, UV light source |
Spectrophotometric Mode | Flat Grating |
Senso | Silicon photodiode array (double row 32 groups) |
Wavelength Range | 400~700nm |
Wavelength Interval | 10nm |
Semiband Width | 10nm |
Measured Reflectance Range | L:0~120; reflectivity:0~200% |
Measuring Aperture | Single Apertures:Φ8mm/Φ10mm |
Specular Component | SCI/SCE |
Color Space | CIE LAB,XYZ,Yxy,LCh,s-RGB,βxy,Munsell(C/2) |
Color Difference Formula | ΔE*ab,ΔE*94,ΔE*cmc(2:1),ΔE*cmc(1:1),ΔE*00, DINΔE99 |
Other Colorimetric Index | WI(ASTM E313,CIE/ISO,AATCC,Hunter), |
YI(ASTM D1925,ASTM 313), | |
Metamerism Index MI, | |
Staining Fastness, Color Fastness, Color Strength, Opacity,Color Card Search | |
Observer Angle | 2°/10° |
Illuminant | D65,A,C,D50,F2(CWF),F7(DLF),F10(TPL5),F11(TL84),F12(TL83/U30) |
Displayed Data | Spectrogram/Values, Samples Chromaticity Values, Color Difference Values/Graph, PASS/FAIL Result, Color Simulation, Color Offset |
Displayed Accuracy | 0.01 |
Measuring Time | About 1.5s (Measure SCI & SCE about 3.2s) |
Repeatability | Chromaticity value: MAV/SCI, within ΔE*ab 0.06 ( When a white calibration plate is measured 30 times at 5 second intervals after white calibration) |
Inter-instrument Error |
MAV/SCI, Within ΔE*ab 0.3 (Average for 12 BCRA Series II color tiles) |
Measurement Mode | Single Measurement, Average Measurement(2-99times) |
Locating Method | Camera Locating,stabilizer cross position |
Dimension | L*W*H=81X71X214mm |
Weight | About 460g |
Battery | Li-ion battery, 6000 measurements within 8 hours |
Illuminant Life Span | 5 years, more than 3 million times measurements |
Display | 3.5-inch TFT color LCD, Capacitive Touch Screen |
Data Port | USB |
Data Storage | Standard 1000 Pcs, Sample 20000 Pcs(One data is able to include SCI/SCE) |
Language | Simplified Chinese, English, Traditional Chinese |
Operating Environment | 0~40℃, 0~85%RH (no condensing), Altitude < 2000m |
Storage Environment | -20~50℃, 0~85%RH (no condensing) |
Standard Accessory | Power Adapter, USB Cable, User Guide, SQCX Software(Download from office website), White and Black Calibration Cavity, Protective Cover, Wrist strap, 8mm flat aperture, 8mm tip aperture |
Optional Accessory | USB Micro Printer, Powder Test Box, |
Notes | Technical parameters are only for reference, subject to the actual sale of the product |